SECTION III. CALIBRATION PROCESS

Calibration Procedure For Igniter Circuit Tester Models 101-5Bf And 101-5Bfg Page Navigation
  1    2  3
img
. . . .
TB 9-6625-1339-24
(9) Remove multimeter from equipment setup.
.
(10) Adjust resistance standard to an initial setting of 0.15
.
(11) Adjust TI OHMS digital dial for an indication of 0.15
(12) Depress and hold TI KEY and adjust resistance standard until TI meter
a
indicates a null. Release KEY. Resistance standard will indicate between 0.13 and 0.17
fter value recorded in (4) above has been added.
(13) Repeat technique of (10) through (12) above, using values listed in table 3 for
models 101-5BF and 101-5BFG. Resistance standard indication plus value recorded in a (4)
above will be within limits specified.
Table 3. Resistance Check, Models 101-5BF and 101-5BFG
Resistance
Resistance standard
Test Instrument
final indication ( )
standard initial
OHMS ADD
OHMS digital
setting ( )
switch setting
dial setting
Min
Max
0.5
0
0.5
0.48
0.52
0.9
0
0.9
0.88
0.92
1.0
0
1.0
0.98
1.02
2.0
0
2.0
1.98
2.02
4.0
0
4.0
3.98
4.02
8.0
0
8.0
7.95
8.05
10
10
0
9.95
10.05
20
20
0
19.95
20.05
25
20
5
24.95
25.05
30
20
10
29.95
30.05
b. Adjustments. No adjustments can be made.
9. Final Procedure
a. Replace shorting bar removed in 7 c above.
b. Deenergize and disconnect all equipment.
c. Annotate and affix DA label/form in accordance with TB 750-25.
4