TB 9-6625-1356-35
Table 1. Calibration Description - Continued.
Test instrument parameters
Performance specifications
Distortion
Model 651B and 652A (SG-763/U)
Range: 10 Hz to 2 MHz
Accuracy: <1%
Range: 2 to 5 MHz
Accuracy: <2%
Range: 5 to 10 MHz
Accuracy: <4%
Models 651A
Range: 10 Hz to 5 MHz
Accuracy: 1%
Approximately 2% at 10 MHz
Output Monitor
Range: 0.9 to 1 V (or 2.8 to 3.2 V)
Accuracy: 2% FS
Model 651B and 652A (SG-763/U) only
Output Monitor Flatness
Model 651B and 652A (SG-763/U) only
Range: 3 and 1 V2
Accuracy: 0.25%
Range: .3 to .3 mV
Accuracy: 0.75%
Range: .1 mV
Accuracy: 1.75%
Frequency Response
Model 651B and 652A (SG-763/U)
Range: 1 to 3 V3
Accuracy: 2%
Frequency: 20 Hz-4 MHz
Accuracy: 1%
Models 651A
Range: 1 to 3 V3
Frequency: 10 to 100 Hz
Accuracy: 3%
Frequency: 100 Hz to 4 MHz
Accuracy: 2%
Frequency: 4 to 10 MHz
Accuracy: 4%
1Procedure
limitation: Certification from -50 to +20 dB only.
2Amplitude
readjusted using expanded scale on output monitor.
3Specification is verified on the 3-V range only.
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286; AN/GSM-287; or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories issued as indicated in paragraph 4 above, and are not listed in this