TB 9-6625-136-24
(18) Press TI TEST button.
0.010 G .
(19) Adjust R14 until TI indicates 1.000 G
(20) Remove resistance standard leads from TI.
(21) Set TI ON/OFF switch to OFF. Remove power cord from TI.
(22) Reassemble TI.
(23) Repeat paragraph 9.
Figure 1. Adjustments.
10. Final Procedure
a. De-energize and disconnect all equipment.
b. Annotate DA Label 163 Limited Calibration: "Not calibrated above 10 G ". Affix in
accordance with TB 750-25.