TB 9-6625-140-24
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Dc voltage:
Fluke, Model 5720A (5720A)
Range:
0 to 1000 V
(p/o MIS-35947); w/ amplifier,
Accuracy: 0.0125%
Fluke 5725A/AR (5725A/AR)
Ac voltage:
Range:
0 to 1000 V
Accuracy: 0.125%
Dc current:
400 A to 10 A
Range:
Accuracy: 0.05%
Ac current:
400 A to 10 A
Range:
Accuracy: 0.05%
Range:
0 to 20 M
Accuracy: 0.0125%
FUNCTION
Range: 0 to 9 MHz
Agilent, Model 33250A (33250A)
Accuracy: 0.0125%
Range: 25 C to 250 C
Azonix, Model A1012
THERMOMETER
(MIS 38958) w/Temperature Probe
Accuracy: 0.25%
Instrulab, Model 4101-10X
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturers' manuals for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
Hz(volts) and COM inputs.
a. Connect calibrator to TI V
b. Set TI function switch mV (mV dc) and the RANGE to 20mV.