TB 9-6625-141-24
(2) Set TI range to the 20 V range and set calibrator output for setting for 6 V. TI will
indicate within the limits in table 3 for the 20 V range.
(3) Repeat step (2) above for the 14 V calibrator output. TI will indicate within the
limits for the 14V output.
(4) Repeat step (2) above for the 19 V calibrator output. TI will indicate within the
limits for the 19 V output. Reverse the calibrator leads for negative 19 V. TI will indicate
within the limits for the negative 19 V output. Return the leads.
(5) Set TI range to the 200 V range and set calibrator output for setting for 190 V. TI
will indicate within the limits for the 200 V range.
(6) Set TI range to the 1000 V range and set calibrator output for setting for 950 V. TI
will indicate within the limits for the 1000 V range.
(7) Set calibrator output to minimum.
Test Instrument
Calibrator
RANGE
Limits
Output
(V dc)
(V dc)
(V dc)
Min
Max
1.9
2000 m
1888 m
1912 m
6
20
5.95
6.05
14
20
13.91
14.09
1
19
20
18.88
19.12
190
200
188.8
191.2
950
1000
943
957
1
Check both (+) and (-) voltages.
b. Adjustments. No adjustments can be made.
a. Performance Check
V
2000 mV.
(1) Set TI function switch to AC volts
(2) Set calibrator for an output of 1.9 V ac at a frequency of 60 Hz. TI will indicate
within limits specified in table 4 for the 2000 mV range.
(3) Repeat technique of (2) above for the 200 and 750 volt ranges. TI will indicate
within limits specified in table 4 for those ranges.
(4) Set calibrator output to minimum.
Test Instrument
Calibrator
Limits
RANGE
Output
(V ac)
(V ac)
(V ac) @ 60 Hz
Min
Max
1.9
2000 mV
1869 m
1931 m
190
200
186.9
193.1
700
750
687
713
b. Adjustments. No adjustments can be made.