TB 9-6625-1493-35
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2
identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-287 or Secondary Reference Calibration Standards Set
NSN 4931-00-621-7878. Alternate items may be used by the calibrating activity. The
items selected must be verified to perform satisfactorily prior to use and must bear evidence
of current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the
standard and TI. Where the four-to-one ratio cannot be met, the actual accuracy of the
equipment selected is shown in parenthesis.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories issued as indicated in paragraph 4 above and are not listed in this
calibration procedure.
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
AC MEASUREMENT
Range: 2.0 mV to 1000 V ac
John Fluke, Model 5790A
STANDARD
Accuracy: 0.0139 %
(5790A)
CALIBRATOR
Dc Voltage:
John Fluke, Model 5720A/CT
Range: 100 mV to 1000 V
(p/o MIS-35947); w/power
Accuracy: 12.5 ppm1,2
amplifier John Fluke, Model
5725 A/CT (5725A/CT)
Ac voltage3,4
Range: 100 mV to 1000 V
.400 & 4.0
9.0
15 & 20
40
100
Accuracy: (in %)
.025
.037
.075
.125
.25
DC REFERENCE
Range: 10 V
John Fluke, Model 732A (732A)
STANDARD
Accuracy: Test report4
DC VOLTAGE DIVIDER
Range: 10:1
ESI, Model RV726 (MIS-10295)
Accuracy:1
(Reference); ESI, Model RV
722 (RV722) (Transfer)
MULTIMETER
Range: 100 mV to 1000 V dc
John Fluke, Model 3458A/E02
(REFERENCE)
Accuracy:4
(MIS-35947/1 Type 2)
POWER SUPPLY
Range: 8.0 to 10 kV
John Fluke, Model 410B/AT
Accuracy: 0.75%
(MIS-10230)
Range: 0 to 1.000000
ESI, Model DT72A (7915908)
Accuracy: Test report3
STANDARD RESISTOR
Range: 10 kΩ
General Radio, Model 1444A
(MIS-10400)
Accuracy: Test report4
Combined accuracy of calibrator (10 V dc) and dc voltage divider is 7.275 ppm.
1
2Combined
accuracy of calibrator (1000 V dc) and multimeter (100 mV and 1.0 V dc) is 25 ppm.
3Combined accuracy of calibrator (1 V, 400 Hz) and ratio transformer for following outputs is: 1 mV, .625%; 10 mV, .075%;
and 100 mV, .0195%.
4Utilize; manufacturer's 24-hour specifications for dc voltage (1 through 1000 V); dc reference standard and standard
resistor are used only to characterize multimeter.