• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table B-53. Calibration Description.
Table B-59. AC Current (=1000 A) (w/50-Turn Coil).Calibrator output

Calibration Procedure For Digital Clamp Meters (General) Page Navigation
  24    25    26    27    28  29  30    31    32    33    34  
img
TB 9-6625-174-24
APPENDIX B
PERFORMANCE TABLES
Extech, Model 38389 Continued
Table B-55. AC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V ac)
(V ac)
(V ac)
Min
Max
3.6 @ 50 Hz
4
3.559
3.641
3.6 @ 60 Hz
4
3.559
3.641
36 @ 50 Hz
40
35.43
36.57
36 @ 60 Hz
40
35.43
36.57
360 @ 50 Hz
400
354.3
365.7
360 @ 60 Hz
400
354.3
365.7
540 @ 50 Hz
600
523
557
Table B-56. DC Current (≤500 A) (w/50-Turn Coil).
Calibrator output
Test instrument
Applied
Range
Limits (A dc)
(A dc)
(A dc)
Min
Max
1.2
400
57.9
62.1
5.4
400
261.6
278.4
7.2
400
347.1
372.9
2.7
600
128
142
6.3
600
301.6
328.4
Table B-57. DC Current (≤1000 A) (w/50-Turn Coil).
Calibrator output
Ammeter calibrator
Test instrument
Applied
Range
Range
Limits (A dc)
(V dc)
(A dc)
(A dc)
Min
Max
1.08
20
600
512
568
Table B-58. AC Current (≤500 A) (w/50-Turn Coil).
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(A ac)
(A ac)
Min
Max
1.2 @ 50 Hz
400
57.6
62.4
1.2 @ 60 Hz
400
57.6
62.4
5.4 @ 50 Hz
400
260.3
279.7
5.4 @ 60 Hz
400
260.3
279.7
7.2 @ 50 Hz
400
346.8
373.2
7.2 @ 60 Hz
400
346.8
373.2
2.7 @ 50 Hz
600
127
143
2.7 @ 60 Hz
600
127
143
6.3 @ 50 Hz
600
299
331
6.3 @ 60 Hz
600
299
331
B-16


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.