TB 9-6625-174-24
APPENDIX B
PERFORMANCE TABLES
B&K Precision, Model 369B Continued
Test instrument
Limits (Hz)
Range
Applied
(Hz)
Min
Max
89.911
90.191
2.5 V rms @ 90 Hz
100
2.5 V rms @ 900 Hz
1k
898.7
901.3
2.5 V rms @ 10 kHz
10 k
8.987 k
9.013 k
2.5 V rms @ 90 kHz
100 k
89.83 k
90.17 k
357.61
362.41
2.5 V rms @ 360 kHz
400 k
Frequencies <100 Hz and >100 kHz may tend to be unstable.
1
Test instrument
Limits (F)
Applied
Range
(μF)
Min
Max
1
0.001
4 nF
0.945 n
1.055 n
40 nF1
0.01
9.81 n
10.19 n
0.1
400 nF
98.1 n
101.9 n
4 μF
1.375 μ
1.425 μ
1.4
40 μF
9.81 μ
10.19 μ
10
Zero TI in Relative Mode before measurement.
1
B-12