Table B-131. AC Voltage.
Table B-138. Calibration Description.

Calibration Procedure For Digital Clamp Meters (General) Page Navigation
  47    48    49    50    51  52  53    54    55    56    57  
img
.
TB 9-6625-174-24
APPENDIX B
PERFORMANCE TABLES
Greenlee, Model CM-950 - Continued
Table B-136. Resistance.
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
190
400
-1.1579 (187.8)
+1.1579 (192.2)
Table B-137. Frequency (≤500 A) (w/Single Conductor).
Calibrator output
Test instrument
Applied
Range
Limits (kHz)
(kHz)
(kHz)
Min
Max
0.050 @ 3 A rms
0.4
0.048
0.052
0.060 @ 3 A rms
0.4
0.058
0.062
0.360 @ 3 A rms
0.4
0.358
0.362