TB 9-6625-174-24
APPENDIX B
PERFORMANCE TABLES
Extech, Model EX810 Continued
Calibrator output
Test instrument
Limits (kHz)
Applied
Range
(Hz)
Min
Max
40 @ 100 V
4.000 k
0.037
0.043
300 @ 50 V
4.000 k
0.293
0.307
3600 @ 5 V
4.00
3.544
3.656
Substitute calibrator for function generator in step 15 a (1).
1
Test instrument
Applied
Range
Limits (F)
(F)
(F)
Min
Max
3n
4n
2.820 n
3.180 n
36 n
40 n
34.00 n
38.00 n
360 n
400 n
348.7 n
371.3 n
3.6 μ
4μ
3.487 μ
3.713 μ
36 μ
40 μ
34.87 μ
37.13 μ
360 μ
400 μ
344.6 μ
375.4 μ
3.6 m
4m
3.230 m
3.970 m
Cool/Heat source
Test instrument
Limits
Applied
Range
Min
Max
0 F
F
-9 F
9 F
1000 F
F
991 F
1039 F
0 C
C
-5 C
5 C
600 C
C
577 C
623 C
B-61