Figure 1. Calibration Label and Adjustment Location.
Table 4. 4410-8 Element Accuracy.

Calibration Procedure For Radio Frequency Power Test Set, An/Urm-213 Page Navigation
  2    3    4    5    6  7  8    9    10    11    12  
img
. . . . . . .
TB 9-6625-177-24
9.
Plug-in Element 4410-8 Calibration
a. Performance Check
(1) Connect equipment as shown in figure 2 below.
Figure 2. Plug-In Element Calibration - Equipment Setup.
(2) Insert plug-in element into TI and rotate until the arrow points in the direction
of the 50 Ω termination.
(3) Set TI range switch to 10 and allow equipment to warm-up for 5 minutes.
(4) Adjust signal generator and RF power amplifier frequency controls for
1000 MHz.
(5) Adjust signal generator and RF power amplifier amplitude controls for a TI
indication of 100 W. RF power meter will indicate within limits in table 4.
(6) Repeat technique of steps (4) and (5) above for remaining settings listed in
table 4. RF power meter will indicate within limits in table 4.