TB 9-6625-1873-50
SECTION I
IDENTIFICATION AND DESCRIPTION
This bulletin provides instructions for the
1. Test Instrument Identification.
75 and TP-93. The manufacturer's manual was used as the prime data source in compiling
instrument" throughout this bulletin.
a. Model Variations. Variations among models are described in text.
b. Time and Technique. The time required for this calibration is approximately 1
hour, using the microwave technique.
2. Calibration Data Card (DA Form 2416). Maintenance forms, records, and reports
which are to be used by calibration personnel at all calibration levels are listed in and
prescribed by TM 38-750.
3. Calibration Description.
Test instrument parameters
and
performance
specifications which pertain to this calibration are listed in table 1.
Table 1. Calibration Description
Test Instrument Parameters
Performance Specifications
VSWR
Less than 1.2 to 1 from 100 to 200 MHz.
Insertion loss
Less than 0.25 db (0.4 db for TP-93) from 1 to 200 MHz.
Impedance
Input, 50 ohms. Output, 75 ohms for TP75 and 93 ohms for TP-93.
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2
identifies the
specific equipment used in
this
calibration procedure. This equipment is issued with secondary transfer calibration
standards set 4931-621-7877 and is to be used in performing this procedure. Alternate
items may be used by the calibrating activity when the equipment listed in table 2 is not
available. The items selected must be verified to perform satisfactorily prior to use and
bear evidence of current calibration. The equipment must meet or exceed the minimum use
specifications listed in table 2. The accuracies listed in table 2 provide a four-to-one
accuracy ratio between the standard and test instrument. Where the four-to-one ratio
cannot be met the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The
accessories listed in
table 3 are issued with
secondary
transfer calibration standards set 4931-621-7877 and are to be used in this calibration
procedure. When necessary, these items may be substituted by equivalent items unless
specifically prohibited.
2