TB 9-6625-1914-35
(4) Repeat (3) above for bands listed in table 8. Noise level will be less than value
listed for each band.
Band
Noise level
(GHz)
(dBm)
.01-2.05
-105
2.60-4.65
-107
4.11-6.15
-105
2.07-6.15
-98
6.17-10.25
-95
4.13-10.25
-93
8.23-14.35
-90
6.19-14.35
-85
10.29-18.45
-80
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
HEWLETT-PACKARD, MODEL 8556( )
41. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 41 and
42 are preparatory to the
calibration
process. Personnel should become familiar with the applicable sections before beginning
the calibration.
b. Items of equipment used in this procedure are referenced within the text by common
WARNING
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.