(6) Carefully adjust amplitudes of function generator and signal generator so that
both signal peaks are 3 dB below LOG REF graticule on display screen. The signals at
140 kHz (2f1) and 180 kHz (2f2) (fig. 17) are second order harmonics. Any second order
intermodulation product (due to analyzer) will occur at 160 kHz (f1 + f2). Any third
order intermodulation products will occur at 50 kHz (2f1-f2) and at 110 kHz (2f2-f1). The
intermodulation products should be below -70 dB graticule line.
(8) Position SCAN WIDTH PER DIVISION switch to 500 Hz and BANDWIDTH
switch to 30 Hz for model 8552B or 50 Hz for model 8552A. Set SCAN WIDTH to 0-
10f.
NOTE
It may be necessary to tune ZERO ADJ control until local
oscillator feed-through is centered at far left graticule line.
(9) Adjust function generator and signal generator controls to set both signal peaks
3 dB below LOG REF graticule on display screen. The signals at 3.4 kHz (2f1) and 4.0
kHz (2f2) are second order harmonics (fig. 17). Any second order intermodulation
product (due to analyzer) will occur at 3.7 kHz (2f1-f2). Any third order intermodulation
product will occur at 1.4 kHz (2f1-f2) and at 2.3 kHz (2f2-f1). All intermodulation
products should be below -60 dB graticule line.
b. Adjustments. No adjustments can be made.
61