TB 9-6625-1924-24
a. Performance Check
(1) Turn FUNCTION switch to +DCV.
(2) Short VOLTS terminal to COM terminal, and OHMS terminal to COM
terminal, using two leads (B2).
(3) Adjust DC ZERO/
control for zero indication on top scale.
(4) Turn FUNCTION switch to OHMS.
If meter does not indicate zero,
perform b below.
(5) Remove shorts connected in (2) above.
b. Adjustments. Adjust A1R5 REG ADJ (fig. 2) while turning FUNCTION switch
between OHMS and +DCV until meter indication is the same in both OHMS and +DCV
positions (R).
a. Performance Check
(1) Connect calibrator to TI VOLTS and COM terminals.
(2) Turn TI FUNCTION switch to +DCV.
(3) Adjust calibrator output for an indication of 1 on TI meter 0 to 1 scale.
If
calibrator does not indicate between 0.098 and 0.102 volt, perform b (1) and (2) below.
(4) Repeat technique of (3) above at TI RANGE switch settings and meter
indications listed in table 3. Calibrator will indicate within limits specified.
Calibrator
Test instrument
indication (V)
RANGE
Meter indication
switch setting
0-1 Scale
0 - 3 Scale
Min
Max
0.3061
.3
---
3
0.294
1.022
1
1
---
0.98
3
---
3
2.94
3.06
10
1
---
9.8
10.2
10
.8
---
7.8
8.2
10
.6
---
5.8
6.2
10
.4
---
3.8
4.2
10
.2
---
1.8
2.2
30
---
3
29.4
30.6
100
1
---
98
102
300
---
3
294
306
1000
1
---
980
1020
1If
indication is out of tolerance, perform b (3) and (4).
2If
indication is out of tolerance, perform b (5) and (6).