TB 9-6625-1956-35
Table 3. Accessories Required
Item
Common name
Description
(part number)
B1
ADAPTER 1
BNC T-type, 2 jacks, 1 plug (MS35173-274C)
B2
ADAPTER 1
BNC jack terminations (MS35184-914)
B3
ADAPTER 1
BNC plug to double banana jacks (7909401)
B4
CABLE 2
30-in., RG-58/U; BNC plug terminations
(7907467)
B5
CABLE
36-in., BNC plug to two alligator clips
(7909410)
B6
CABLE
Thermistor mount cable (8120-1082)
1Two required
2 Four required.
SECTION III
CALIBRATION PROCESS
6.
Preliminary Instructions
WARNING
HIGH VOLTAGE is used or exposed during
the performance of this calibration. DEATH
should become familiar with the entire bulletin before
ON CONTACT may result if personnel fail to
beginning the calibration.
observe safety precautions.
b. Items of equipment used in this procedure are
referenced within the text by common name and item
a. Set TI 115-230 V switch (rear panel) to 115 V.
the identification of equipment referenced by item
numbers prefixed with A, see table 2, and for prefix B,
adjust for a 115-V ac output.
d. Position TI controls as listed in (1) through (3)
c. Unless otherwise specified, verify the result of
below:
each test and, whenever the test requirement is not met,
take corrective action before continuing with the
(1) ZERO/TEST switch to ZERO.
calibration. Adjustments required to calibrate the TI are
included in this procedure. Additional maintenance
(2) FUNCTION switch to 200Q.
information is contained in the manufacturer's manual for
this- TI.
(3) POWER (mW) switch to .01.
d. Unless otherwise specified, all controls and
control settings refer to the TI.
e. Energize TI and allow 30 minutes for warmup and
stabilization
7. Equipment Setup
4