TB 9-6625-1958-24
(5) Set measuring receiver to measure volts at 500 MHz. If measuring receiver
indication is less than 265.1 mV, perform b (2) through (14) below.
(6) D i s c o n n e c t s e n s o r m o d u l e f r o m T I M A R K E R O U T a n d c o n n e c t t o T I
1 n S ONLY.
(7) Press 1nS pushbutton.
(8) Set measuring receiver to measure volts at 1000 MHz. If measuring receiver
indication is less than 70.7 mV, perform b (8) through (14) below.
b. Adjustments.
(1) Adjust C450, C465, and C467 (fig. 2) for maximum measuring receiver
indication (R).
(2) Adjust C498 and R482 (fig. 2) for maximum measuring receiver indication (R).
(3) Adjust C504 and C505 (fig. 2) for maximum measuring receiver indication (R).
(4) Adjust C510, C511, C512, and C513 (fig. 2) (order of adjustments as shown by
arrow) for maximum measuring receiver indication (R).
(5) D i s c o n n e c t s e n s o r m o d u l e f r o m T I M A R K E R O U T a n d c o n n e c t t o T I
1 n S ONLY.
(6) Press 1nS pushbutton.
(7) Set measuring receiver to measure volts at 1000 MHz.
7