TB 9-6625-1973-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Frequency Counter, Agilent Models 5326A, 5326B, and 5326C. The
manufacturer's manual was used as the prime data source in compiling these instructions.
The equipment being calibrated will be referred to as the TI (test instrument) throughout
this bulletin.
a. Model Variations. Variations among models are described in the text.
b. Time and Technique. The time required for this calibration is approximately 2
hours per instrument, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument
parameters
Performance specifications
Time base
10 MHz
Frequency
< 3 parts in 107/month (5 parts in 1010/day for Model 5326A with Option 011)
Aging Rate
1 part in 107 (5 parts in 109 for Model 5326A with Option 011) for 10% line
Stability
voltage change
Frequency
A and B
0 to 50 MHz
Sensitivity
100 mV
Input zero
Less than + 1 mV dc at A and B INPUT jacks for X10 and X100 positions of
attenuator (Option 004 only)
Frequency C1
Range
0 to 50 MHz
5 mV2
Sensitivity
Dc voltage3
Ranges
10, 100, and 1000 V dc
Accuracy
10 and 100 V ranges, (0.04% of reading + 0.02% of range)
1 count; 1000V
range, (0.08% of reading + 0.02% of range) 1 count
1Not
available on Model 5326C.
2Test
instruments with serial number prefix 1128A and below, 0 to 30 MHz, 50 mV; 30 MHz to 50 MHz, 100 mV.
3Available on Model 5326B only.