TB 9-6625-1976-35
19. Power Supply
(Model 5215A)
a. Performance Check
NOTE
Do not perform power supply checks if all other parameters are
within tolerance.
(1) Connect multimeter between TP11 +15V and TP 19 GRD (fig. 3). If multimeter
does not indicate between 14.95 and 15.05 V, perform b below.
(2) Connect multimeter between TP 12 - 15 V and TP 19 GND (fig. 3). Multimeter
will indicate between -14.65 and -15.35 V.
b. Adjustments.
Adjust R82 +15V (fig. 3)
for a
multimeter indication between +14.95
and 15.05 V dc (R).
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.