TB 9-6625-1978-35
(8) Adjust signal generator frequency from 10 to 135 MHz. TI will display a stable
count of applied frequency.
(9) Disconnect signal generator and 50Ω feedthrough termination from TI.
(10) Set BAND SELECT switch to 10 MHz - 300 MHz.
(11) Connect signal generator RF Output to TI 10 MHz - 300 MHz 50 OHM input.
(12) Set signal generator frequency to 10 MHz and level output to 20 dBm. TI will
display a stable count of applied frequency.
(13) Adjust signal generator frequency from 10 to 300 MHz. TI will display a stable
count of applied frequency.
b. Adjustments. No adjustments can be made.
15. Band II Input
Sensitivity
a. Performance Check
(1) Set BAND SELECT switch to 100 MHz - 850 MHz.
(2) Connect signal generator RF Output to TI 100 MHz - 850 MHz 50 OHM input.
(3) Set signal generator frequency to 100 MHz and level output to 50 dBm.
(4) Slowly increase signal generator level output until TI displays a stable count of
(6) Set signal generator frequency to 850 MHz and level output to 50 dBm.
(7) Slowly increase signal generator level output until TI displays a stable count of
was previously made, perform b below.
b. Adjustments
(1) Set signal generator frequency to 100 MHz and level output to -15 dBm.
(2) Adjust A109R41 (fig. 1) until TI displays a stable count of applied frequency (R).
(3) Slowly reduce signal generator level output until TI display becomes erratic.
(4) Readjust Al09R41 (fig. 1) until TI indication drops to all zeros.
16. Band III Input
Sensitivity
a. Performance Check
(1) Set BAND SELECT switch to 825 MHz - 18 GHz.
(2) Connect signal generator RF Output to TI 825 MHz - 18 GHz 50 OHM input
(3) Set signal generator frequency to 825 MHz and level output to 50 dBm.
9