TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Amprobe Instruments, Model 33XR-A - Continued
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
190
400
-1.211% (187.7)
+1.211% (192.3)
19 k
40 k
-1.211% (18.77 k)
+1.211% (19.23 k)
1.9 M
4M
-1.421% (1.873 M)
+1.421% (1.927 M)
Test instrument
output
Applied
Range
Limits (F)
(F)
(F)
Min
Max
3 μ
4μ
2.840 μ
3.160 μ
36 μ
40 μ
34.15 μ
37.85 μ
360 μ
400 μ
341.5 μ
378.5 μ
3600 μ
4000 μ
μ
μ
3405
3795
Test instrument
Applied
Limits (Hz)
(V ac)
Min
Max
1.6 @ 3.6 kHz
3.593 k
3.607 k
1.6 @ 36 kHz
35.93 k
36.07 k
1.6 @ 360 k
359.3 k
360.7 k
1.6 @ 3.6 MHz
3.593 M
3.607 M
2.1 @ 20 MHz
19.95 M
19.95 M
Table B-86. Temperature.
Cool/Heat source
Test instrument
Applied
Limits ()
()
Range
Min
Max
0C
C
-4 C
4C
180 C
C
175 C
185 C
500 C
C
483 C
517 C
F1
32 F
23 F
41 F
360 F
F
350 F
370 F
900 F
F
869 F
931 F
Set TI Function/Range switch to F.
1