• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table B-72. Calibration Description.
Table B-75. DC Current.

Calibration Proceudre For Digital Multimeters (General) Page Navigation
  35    36    37    38    39  40  41    42    43    44    45  
img
TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Beckman Instruments, Model 330 - Continued
Table B-74. AC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V ac)
(V ac)
(V ac)
Min
Max
190 m @ 50 Hz
200 m
188.6 m
191.4 m
190 m @ 400 Hz
200 m
188.6 m
191.4 m
190 m @ 1.5 kHz
200 m
188.6 m
191.4 m
190 m @ 3 kHz
200 m
187.7 m
192.3 m
190 m @ 8 kHz
200 m
186.6 m
193.4 m
190 m @ 15 kHz
200 m
185.5 m
194.5 m
180 m @ 30 kHz
200 m
169.4 m
190.6 m
1.9 @ 50 Hz
2
1.886
1.914
1.9 @ 400 Hz
2
1.886
1.914
1.9 @ 1.5 kHz
2
1.886
1.914
1.9 @ 3 kHz
2
1.887
1.923
1.9 @ 8 kHz
2
1.866
1.934
1.9 @ 15 kHz
2
1.855
1.945
1.8 @ 30 kHz
2
1.694
1.906
19 @ 50 Hz
20
18.86
19.14
19 @ 400 Hz
20
18.86
19.14
19 @ 1.5 kHz
20
18.86
19.14
19 @ 3 kHz
20
18.77
19.23
19 @ 8 kHz
20
18.66
19.34
19 @ 15 kHz
20
18.55
19.45
18 @ 30 kHz
20
16.94
19.06
190 @ 50 Hz
200
188.6
191.4
190 @ 400 Hz
200
188.6
191.4
190 @ 1.5 kHz
200
188.6
191.4
190 @ 3 kHz
200
187.7
192.3
190 @ 8 kHz
200
186.6
193.4
190 @ 15 kHz
200
185.5
194.5
900 @ 50 Hz
1000
892
908
900 @ 400 Hz
1000
892
908
900 @ 1.5 kHz
1000
892
908
900 @ 3 kHz
1000
887
913
900 @ 8 kHz
1000
881
919
900 @ 15 kHz
1000
875
925
B-30