• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table B-62. AC Voltage.
Table B-66. Calibration Description.

Calibration Proceudre For Digital Multimeters (General) Page Navigation
  30    31    32    33    34  35  36    37    38    39    40  
img
TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Beckman Instruments, Model 3010 - Continued
Table B-64. AC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(A ac)
(A ac)
Min
Max
190 μ @ 45 Hz
200 μ
186.8  μ
193.2  μ
190 μ @ 2 kHz
200 μ
186.8  μ
193.2  μ
1.9 m @ 45 Hz
2m
1.868 m
1.932 m
1.9 m @ 2 kHz
2m
1.868 m
1.932 m
19 m @ 45 Hz
20 m
18.68 m
19.32 m
19 m @ 2 kHz
20 m
18.68 m
19.32 m
190 m @ 45 Hz
200 m
186.8  m
193.2  m
190 m @ 2 kHz
200 m
186.8  m
193.2  m
1.9 @ 45 Hz
2
1.868
1.932
1.9 @ 2 kHz
2
1.868
1.932
9 @ 45 Hz
10
8.79
9.21
9 @ 2 kHz
10
8.79
9.21
Table B-65. Resistance.
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
190
200
-0.58% (188.9)
+0.58% (191.1)
1.9 k
2k
-0.58% (1.889 k)
+0.58% (1.911 k)
19 k
20 k
-0.58% (18.89 k)
+0.58% (19.11 k)
190 k
200 k
-0.58% (188.9 k)
+0.58% (191.1 k)
1.9 M
2M
-0.58% (1.889 M)
+0.58% (1.911 M)
19  M
20 M
-1.58% (18.70 M)
+1.58% (19.30 M)
B-25