• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table B-52. Calibration Description.
Table B-57. Resistance.

Calibration Proceudre For Digital Multimeters (General) Page Navigation
  26    27    28    29    30  31  32    33    34    35    36  
img
TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Wavetek, Model 25XT - Continued
Table B-54. AC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V ac)
(V ac)
(V ac)
Min
Max
180 m @ 60 Hz
200 m
177.8  m
182.2  m
180 m @ 400 Hz
200 m
177.8  m
182.2  m
180 m @ 1 kHz
200 m
176.9  m
183.1  m
1.8 @ 60 Hz
2
1.778
1.822
1.8 @ 400 Hz
2
1.778
1.822
1.8 @ 1 kHz
2
1.769
1.831
18 @ 60 Hz
20
17.78
18.22
18 @ 400 Hz
20
17.78
18.22
18 @ 1 kHz
20
17.69
18.31
180 @ 60 Hz
200
177.8
182.2
180 @ 400 Hz
200
177.8
182.2
180 @ 1 kHz
200
176.9
183.1
675 @ 60 Hz
750
664
686
675 @ 400 Hz
750
664
686
675 @ 1 kHz
750
661
689
Table B-55. DC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A dc)
(A dc)
(A dc)
Min
Max
180  μ
200 μ
178.1  μ
181.9  μ
1.8 m
2m
1.781 m
1.819 m
18  m
20 m
17.81 m
18.19 m
90  m
100 m
89.0  m
91.0  m
101
9
8.79
9.21
Move test lead from TI mA Cx terminal to TI 10 A terminal.
1
Table B-56. AC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(A ac)
(A ac)
Min
Max
180 μ @ 60 Hz
200 μ
176.9 μ
183.1  μ
180 μ @ 400 Hz
200 μ
176.9 μ
183.1  μ
1.8 m @ 60 Hz
2m
1.769 m
1.831 m
1.8 m @ 400 Hz
2m
1.769 m
1.831 m
18 m @ 60 Hz
20 m
17.69 m
18.31 m
18 m @ 400 Hz
20 m
17.69 m
18.31 m
90 m @ 60 Hz
100 m
88.2  m
91.8  m
90 m @ 400 Hz
100 m
88.2  m
91.8  m
101
9 @ 60 Hz
8.73
9.27
9 @ 400 Hz
10
8.73
9.27
B-21