• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table B-48. AC Current.
Table B-52. Calibration Description.

Calibration Proceudre For Digital Multimeters (General) Page Navigation
  24    25    26    27    28  29  30    31    32    33    34  
img
TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Wavetek, Model 2030 - Continued
Table B-51. Frequency.
Function generator output
Test instrument
Applied
Limits (Hz)
(V ac)
Min
Max
1 V @ 1 kHz
0.997 k
1.003 k
1 V @ 10 kHz
9.987 k
10.013 k
1 V @ 50 kHz
49.93  k
50.07  k
1 V @ 100 kHz
99.88  k
100.12  k
1 V @ 1 MHz
0.9988 M
1.0012 M
B-19


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.