• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table B-54. AC Voltage.
Table B-60. Calibration Description.

Calibration Proceudre For Digital Multimeters (General) Page Navigation
  27    28    29    30    31  32  33    34    35    36    37  
img
TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Wavetek, Model 25XT - Continued
Table B-57. Resistance.
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
190
200
-1.21% (187.7)
+1.21% (192.3)
1.9 k
2k
-1.21% (1.877 k)
+1.21% (1.923 k)
19  k
20 k
-1.21% (18.77 k)
+1.21% (19.23 k)
190  k
200 k
-1.21% (187.7 k)
+1.21% (192.3 k)
1.9 M
2M
-1.21% (1.877 M)
+1.21% (1.923 M)
19  M
20 M
-2.21% (18.58 M)
+2.21% (19.42 M)
Table B-58. Capacitance.
Capacitance standard
Test instrument
output
Applied
Range
Limits (F)
(F)
(F)
Min
Max
0.0001 μ
200 p
97.5 p
102.5  p
0.001 μ
2n
0.976 n
1.024 n
0.01  μ
20 n
9.76 n
10.24 n
μ
0.1
200 n
97.6  n
102.4  n
μ
2μ
0.976 μ
1.024 μ
1
μ
20 μ
9.76 μ
10.24 μ
10
μ
200 μ
97.6  μ
102.4  μ
100
2000 μ
1
m
0.976 m
1.024 m
10
m
20 m
9.66  m
10.30 m
Table B-59. Frequency.
Function generator output
Test instrument
Applied
Limits (Hz)
(V ac)
Min
Max
2 V @ 20 Hz
17
23
2 V @ 1 kHz
0.996 k
1.004 k
2 V @ 10 kHz
9.96  k
10.04  k
2 V @ 38 kHz
37.7
k
38.3  k
B-22