Table B-66. Calibration Description.
Table B-71. Resistance.

Calibration Proceudre For Digital Multimeters (General) Page Navigation
  32    33    34    35    36  37  38    39    40    41    42  
img
TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Beckman Instruments, Model 320B - Continued
Table B-68. AC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V ac)
(V ac)
(V ac)
Min
Max
190 m @ 45 Hz
200 m
188.3  m
191.7  m
190 m @ 1.9 kHz
200 m
188.3  m
191.7  m
190 m @ 5 kHz
200 m
186.6  m
193.4  m
1.9 @ 45 Hz
2
1.883
1.917
1.9 @ 1.9 kHz
2
1.883
1.917
1.9 @ 5 kHz
2
1.866
1.934
19 @ 45 Hz
20
18.83
19.17
19 @ 1.9 kHz
20
18.83
19.17
19 @ 5 kHz
20
18.66
19.34
190 @ 45 Hz
200
188.3
191.7
900 @ 45 Hz
1000
890
910
900 @ 1 kHz
1000
881
919
Table B-69. DC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A dc)
(A dc)
(A dc)
Min
Max
190  μ
200 μ
189.2  μ
190.8  μ
1.9 m
2m
1.892 m
1.908 m
19  m
20 m
18.92 m
19.08 m
190  m
200 m
189.2  m
190.8  m
1.9
2
1.892
1.908
9
10
8.85
9.15
Table B-70. AC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(A ac)
(A ac)
Min
Max
190 μ @ 45 Hz
200 μ
188.0  μ
192.0 μ
190 μ @ 2 kHz
200 μ
188.0  μ
192.0 μ
1.9 m @ 45 Hz
2m
1.880 m
1.920 m
1.9 m @ 2 kHz
2m
1.880 m
1.920 m
19 m @ 45 Hz
20 m
18.80 m
19.20 m
19 m @ 2 kHz
20 m
18.80 m
19.20 m
190 m @ 45 Hz
200 m
188.0  m
192.0  m
190 m @ 2 kHz
200 m
188.0  m
192.0  m
1.9 @ 45 Hz
2
1.880
1.920
1.9 @ 2 kHz
2
1.880
1.920
9 @ 45 Hz
10
8.83
9.17
9 @ 2 kHz
10
8.83
9.17