TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Beckman Instruments, Model 320B - Continued
Table B-68. AC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V ac)
(V ac)
(V ac)
Min
Max
190 m @ 45 Hz
200 m
188.3 m
191.7 m
190 m @ 1.9 kHz
200 m
188.3 m
191.7 m
190 m @ 5 kHz
200 m
186.6 m
193.4 m
1.9 @ 45 Hz
2
1.883
1.917
1.9 @ 1.9 kHz
2
1.883
1.917
1.9 @ 5 kHz
2
1.866
1.934
19 @ 45 Hz
20
18.83
19.17
19 @ 1.9 kHz
20
18.83
19.17
19 @ 5 kHz
20
18.66
19.34
190 @ 45 Hz
200
188.3
191.7
900 @ 45 Hz
1000
890
910
900 @ 1 kHz
1000
881
919
Table B-69. DC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A dc)
(A dc)
(A dc)
Min
Max
190 μ
200 μ
189.2 μ
190.8 μ
1.9 m
2m
1.892 m
1.908 m
19 m
20 m
18.92 m
19.08 m
190 m
200 m
189.2 m
190.8 m
1.9
2
1.892
1.908
9
10
8.85
9.15
Table B-70. AC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(A ac)
(A ac)
Min
Max
190 μ @ 45 Hz
200 μ
188.0 μ
192.0 μ
190 μ @ 2 kHz
200 μ
188.0 μ
192.0 μ
1.9 m @ 45 Hz
2m
1.880 m
1.920 m
1.9 m @ 2 kHz
2m
1.880 m
1.920 m
19 m @ 45 Hz
20 m
18.80 m
19.20 m
19 m @ 2 kHz
20 m
18.80 m
19.20 m
190 m @ 45 Hz
200 m
188.0 m
192.0 m
190 m @ 2 kHz
200 m
188.0 m
192.0 m
1.9 @ 45 Hz
2
1.880
1.920
1.9 @ 2 kHz
2
1.880
1.920
9 @ 45 Hz
10
8.83
9.17
9 @ 2 kHz
10
8.83
9.17
B-27