TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Wavetek, Model 2030 - Continued
Table B-48. AC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(A ac)
(A ac)
Min
Max
900 μ @ 60 Hz
1000 μ
883.9 μ
916.1 μ
900 μ @ 400 Hz
1000 μ
883.9 μ
916.1 μ
900 μ @ 1 kHz
1000 μ
883.9 μ
916.1 μ
9 m @ 60 Hz
10 m
8.839 m
9.161 m
9 m @ 400 Hz
10 m
8.839 m
9.161 m
9 m @ 1 kHz
10 m
8.839 m
9.161 m
90 m @ 60 Hz
100 m
88.42 m
91.58 m
90 m @ 400 Hz
100 m
88.42 m
91.58 m
90 m @ 1 kHz
100 m
88.42 m
91.58 m
1000 m1
900 m @ 60 Hz
884.2 m
915.8 m
1
900 m @ 400 Hz
1000 m
884.2 m
915.8 m
900 m @ 1 kHz
1000 m
884.2 m
915.8 m
9 @ 60 Hz
10
8.844
9.156
9 @ 400 Hz
10
8.844
9.156
9 @ 1 kHz
10
8.844
9.156
10 @ 60 Hz
20
9.80
10.20
10 @ 400 Hz
20
9.80
10.20
10 @ 1kHz
20
9.80
10.20
Move lead from TI 100 mA terminal to TI 10 A terminal.
1
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
100
100
-0.54% (99.46)
+0.54% (100.54)
1k
1000
-0.34% (996.6)
+0.34% (1003.4)
10 k
10 k
-0.34% (9.966 k)
+0.34% (10.034 k)
100 k
100 k
-0.34% (99.66 k)
+0.34% (100.34 k)
1M
1000 k
-0.34% (996.6 k)
+0.34% (1003.4 k)
10 M
10 M
-0.74% (9.926 M)
+0.74% (10.074 M)
19 M
20 M
-1.221% (18.768 M)
+1.221% (19.232 M)
Test instrument
output
Applied
Range
Limits (F)
(F)
(F)
Min
Max
0.01 μ
20 n
9.5
n
10.5
n
0.1 μ
200 n
97.7
n
102.3
n
μ
2μ
0.9798 μ
1.0202 μ
1
μ
20 μ
9.798 μ
10.202 μ
10
μ
200 μ
102.97 μ
103.03 μ
100
1
m
2m
0.9692 m
1.0308 m
B-18