TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Wavetek, Model 2020 - Continued
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
1.9 M
10 M
-1.737% (1.867 M)
+1.737% (1.933 M)
19 M
20 M
-2.021% (18.616 M)
+2.021% (19.384 M)
Calibrator output
Test instrument
Limits (Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
90
100
89.33
90.67
900
1000
895.1
904.9
9k
10 k
8.951 k
9.049 k
90 k
100 k
89.51 k
90.49 k
900 k
1000 k
895.1 k
904.9 k
Test instrument
output
Applied
Range
Limits (F)
(F)
(F)
Min
Max
0.01 μ
20 n
9.5
n
10.5
n
0.1 μ
200 n
97.7
n
102.3
n
μ
2μ
0.9798 μ
1.0202 μ
1
μ
20 μ
9.798 μ
10.202 μ
10
μ
200 μ
102.97 μ
103.03 μ
100
1
m
2m
0.9692 m
1.0308 m
Test instrument
output
Applied
Range
Limits (Hz)
(V ac)
(Hz)
Min
Max
1 @ 2 kHz
19.999 k
1.996 k
2.004 k
1 @ 19 kHz
19.999 k
18.979 k
19.021 k
1 @ 190 kHz
199.99 k
189.79 k
190.21 k
1 @ 1.9 MHz
1.9999 M
1.8979 M
1.9021 M
B-15