TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Wavetek, Model 2015 - Continued
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
190
400
-1.000% (188.1)
+1.000% (191.9)
1.9 k
4k
-0.789% (1.885 k)
+0.789% (1.915 k)
19 k
40 k
-0.789% (18.85 k)
+0.789% (19.15 k)
190 k
400 k
-0.789% (188.5 k)
+0.789% (191.5 k)
1.9 M
4M
-1.211% (1.877 M)
+1.211% (1.923 M)
19 M
40 M
-1.737% (18.67 M)
+1.737% (19.33 M)
Test instrument
output
Applied
Range
Limits (F)
(F)
(F)
Min
Max
0.001 μ
4n
0.950 n
1.050 n
0.01 μ
40 n
9.86 n
10.14 n
0.1 μ
400 n
98.6 n
101.4 n
μ
4μ
0.986 μ
1.014 μ
1
μ
40 μ
9.86 μ
10.14 μ
10
Test instrument
output
Applied
Range
Limits (Hz)
(V ac)
(Hz)
Min
Max
40 m @ 20 Hz
100
19.88
20.12
40 m @ 900 Hz
1k
898.7
901.3
40 m @ 9 kHz
10 k
8.987 k
9.013 k
400 m @ 90 kHz
100 k
89.83 k
90.17 k
400 m @ 380 kHz
400 k
377.6 k
382.4 k
B-12