Table B-21. AC Voltage.
Table B-27. Calibration Description.

Calibration Proceudre For Digital Multimeters (General) Page Navigation
  14    15    16    17    18  19  20    21    22    23    24  
img
TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Wavetek, Model 16XL - Continued
Table B-25. Capacitance.
Capacitance standard
Test instrument
output
Applied
Range
Limits (F)
(F)
(F)
Min
Max
0.001 μ
2n
0.961 n
1.039 n
0.01 μ
20 n
9.61 n
10.39 n
0.1  μ
200 n
96.1  n
103.9  n
μ
2μ
0.961 μ
1.039 μ
1
μ
20 μ
9.61 μ
10.39 μ
10
Table B-26. Frequency.
Function generator
Test instrument
output
Applied
Range
Limits (Hz)
(V ac)
(Hz)
Min
Max
1 @ 100 Hz
2k
98
102
1 @ 10 kHz
20 k
9.97 k
10.03 k
1 @ 180 kHz
200 k
179.6  k
180.4  k
1 @ 1.8 MHz
2M
1.796 M
1.804 M
1 @ 13.5 MHz
15 M
13.47 M
13.53 M
B-9