• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table B-27. Calibration Description.
Table B-32. Resistance.

Calibration Proceudre For Digital Multimeters (General) Page Navigation
  16    17    18    19    20  21  22    23    24    25    26  
img
TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Wavetek, Model 2015 - Continued
Table B-29. AC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V ac)
(V ac)
(V ac)
Min
Max
360 m @ 60 Hz
400 m
355.9  m
364.1  m
3.6 @ 60 Hz
4
3.568
3.632
3.6 @ 400 Hz
4
3.568
3.632
3.6 @ 1kHz
4
3.552
3.648
36 @ 60 Hz
40
35.68
36.32
36 @ 400 Hz
40
35.68
36.32
36 @ 1 kHz
40
35.59
36.41
360 @ 60 Hz
400
356.8
363.2
360 @ 400 Hz
400
356.8
363.2
360 @ 1 kHz
400
355.9
364.1
675 @ 60 Hz
750
667
683
675 @ 400 Hz
750
667
683
675 @ 1 kHz
750
662
688
Table B-30. DC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A dc)
(A dc)
(A dc)
Min
Max
3.6 m
4m
3.569 m
3.631 m
36  m
40 m
35.69 m
36.31 m
360  m
400 m
365.9  m
363.1  m
201
10
9.84
10.16
Move lead from TI 400 mA terminal to TI 20 A terminal.
1
Table B-31. AC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(A ac)
(A ac)
Min
Max
3.6 m @ 60 Hz
4m
3.553 m
3.647 m
3.6 m @ 1 kHz
4m
3.553 m
3.647 m
36 m @ 60 Hz
40 m
35.53 m
36.47 m
36 m @ 1 kHz
40 m
35.53 m
36.47 m
360 m @ 60 Hz
400 m
355.3  m
364.7  m
360 m @ 1 kHz
400 m
355.3  m
364.7  m
10 @ 60 Hz
20
9.80
10.20
10 @ 1 kHz
20
9.80
10.20
Move lead from TI 400 mA terminal to TI 20 A terminal.
1
B-11


Privacy Statement
Press Release
Contact