TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Wavetek, Model 15XL Continued
Table B-10. DC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A dc)
(A dc)
(A dc)
Min
Max
190 μ
200 μ
188.0 μ
192.0 μ
1.9 m
2m
1.880 m
1.920 m
19 m
20 m
18.80 m
19.20 m
190 m
200 m
188.0 m
192.0 m
101
9
8.79
9.21
Move test lead from TI 200 mA terminal to 10 A terminal.
1
Table B-11. AC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(A ac)
(A ac)
Min
Max
190 μ @ 50 Hz
200 μ
186.7 μ
193.3 μ
190 μ @ 500 Hz
200 μ
186.7 μ
193.3 μ
1.9 m @ 50 Hz
2m
1.867 m
1.933 m
1.9 m @ 500 Hz
2m
1.867 m
1.933 m
19 m @ 50 Hz
20 m
18.67 m
19.33 m
19 m @ 500 Hz
20 m
18.67 m
19.33 m
190 m @ 50 Hz
200 m
186.7 m
193.3 m
190 m @ 50 Hz
200 m
186.7 m
193.3 m
101
9 @ 50 Hz
8.73
9.27
9 @ 50 Hz
10
8.73
9.27
Move lead from TI 200 mA terminal to 10 A terminal.
1
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
190
200
-1.158% (187.8)
+1.158% (192.2)
1.9 k
2k
-1.05% (1.880 k)
+1.05% (1.920 k)
19 k
20 k
-1.05% (18.80 k)
+1.05% (19.20 k)
190 k
200 k
-1.05% (188.0 k)
+1.05% (192.0 k)
1.9 M
2000 k
-1.05% (1880 k)
+1.05% (1920 k)
19 M
20 M
-3.05% (18.42 M)
+3.05% (19.58 M)
-15.00% (85 M)1
+15.00% (115 M)1
100 M
2000 M
Subtract 10 MΩ from TI displayed value. Result must be within specified limits.
1
B-4