Table B-128. Calibration Description.
Table B-136. DC Voltage.

Calibration Proceudre For Digital Multimeters (General) Page Navigation
  58    59    60    61    62  63  64    65    66    67    68  
img
TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Beckman Instruments, Model DM25 - Continued
Table B-132. AC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(A ac)
(A ac)
Min
Max
190 μ @ 400 Hz
200 μ
186.2 μ
193.8 μ
19 m @ 400 Hz
20 m
18.62 m
19.38 m
190 m @ 400 Hz
200 m
186.2 m
193.8 m
9 @ 400 Hz
10
8.69
9.31
Move TI lead to TI 10 A input.
1
Table B-133. Resistance.
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
100
200/HI
-1.4% (98.6)
+1.4% (101.4)
1k
2 k/HI
-0.9% (0.991 k)
+0.9% (1.009 k)
1k
2 k/LO
-1.4% (0.986 k)
+1.4% (1.014 k)
10 k
20 k/HI
-0.9% (9.91 k)
+1.4% (10.09 k)
10 k
20 k/LO
-1.4% (9.86 k)
+1.4% (10.14 k)
100 k
200 k/HI
-0.9% (99.1 k)
+0.9% (100.9 k)
100 k
200 k/LO
-1.4% (98.60 k)
+1.4% (101.4 k)
1M
2 M/HI
-0.9% (0.991 M)
+0.9% (1.009 M)
1M
2 M/LO
-1.4% (0.986 M)
+1.4% (1.014 M)
10 M
20 M/HI
-2.4% (9.76 M)
+2.4% (10.24 M)
10 M
20 M/LO
-3.4% (9.96 M)
+3.4% (10.34 M)
Table B-134. Capacitance.
Capacitance standard
Test instrument
output
Applied
Range
Limits (F)
(F)
(F)
Min
Max
0.001 μ
2n
0.96 n
1.04 n
0.01 μ
20 n
9.6  n
10.4  n
0.1  μ
200 n
96.0  n
104
n
1.0  μ
2μ
0.960 μ
1.040 μ
μ
20 μ
9.6  μ
10.4  μ
10
B-53