Table B-138. DC Current.
Table B-143. Calibration Description.

Calibration Proceudre For Digital Multimeters (General) Page Navigation
  61    62    63    64    65  66  67    68    69    70    71  
img
TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Beckman Instruments, Model DM25XL - Continued
Table B-142. Frequency.
Function generator output
Test instrument
Applied
Range
Limits (Hz)
(Hz)
(Hz)
Min
Max
1 V @ 1 kHz
2k
0.989 k
1.011 k
1 V @ 10 kHz
20 k
9.89 k
10.11 k
1 V @ 100 kHz
200 k
98.9  k
101.1  k
1 V @ 1 MHz
2M
0.989 M
1.011 M
1 V @ 10 MHz
20 M
9.89 M
10.11 M