Table B-218. AC Voltage.

Calibration Proceudre For Digital Multimeters (General) Page Navigation
  91    92    93    94    95  96
img
TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Beckman Instruments, Model TECH310 - Continued
Table B-220. AC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(A ac)
(A ac)
Min
Max
190 μ @ 45 Hz
200 μ
186.8  μ
193.2 μ
190 μ @ 390 Hz
200 μ
186.8  μ
193.2 μ
190 μ @ 2 kHz
200 μ
186.7  μ
193.3 μ
1.9 m @ 45 Hz
2m
1.868 m
1.932 m
1.9 m @ 390 Hz
2m
1.868 m
1.932 m
1.9 m @ 2 kHz
2m
1.867 m
1.933 m
19 m @ 45 Hz
20 m
18.68 m
19.32 m
19 m @ 390 Hz
20 m
18.68 m
19.32 m
19 m @ 2 kHz
20 m
18.67 m
19.33 m
1.9 @ 45 Hz
2
1.868
1.932
1.9 @ 390 Hz
2
1.868
1.932
1.9 @ 2 kHz
2
1.867
1.933
9 @ 45 Hz
10
8.79
9.21
9 @ 400 Hz
10
8.79
9.21
Table B-221. Resistance.
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
190
200
-0.789% (188.9)
+0789% (191.1)
1.9 k
2k
-0.789% (1.889 k)
+0.789% (1.911 k)
19  k
20 k
-0.789% (18.89 k)
+0.789% (19.11 k)
190  k
200 k
-0.789% (188.9 k)
+0.789% (191.1 k)
1.9 M
2M
-0.789% (1.889 M)
+0.789% (1.911 M)
19  M
20 M
-1.579% (18.70 M)
+1.579% (19.30 M)