TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Wavetek, Model RMS225A Continued
Table B-213. DC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A dc)
(A dc)
(A dc)
Min
Max
9m
10 m
8.924 m
9.076 m
30 m
40 m
29.69 m
30.31 m
9
10
8.924
9.076
Table B-214. AC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(A ac)
(A ac)
Min
Max
9 m @ 50 Hz
10 m
8.834 m
9.166 m
9 m @ 1 kHz
10 m
8.834 m
9.166 m
30 m @ 50 Hz
40 m
29.39 m
30.61 m
30 m @ 1 kHz
40 m
29.39 m
30.61 m
9 @ 50 Hz
10
8.834
9.166
9 @ 1 kHz
10
8.834
9.166
Calibrator output
Test instrument
Limits (Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
900
1k
895.4
904.6
9k
10 k
8.954 k
9.046 k
90 k
100 k
89.54 k
90.46 k
900 k
1M
895.4 k
904.6 k
9M
10 M
8.908 M
9.092 M
36 M
40 M
35.44 M
36.56 M
B-83