• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table B-195. Calibration Description.
Table B-200. Resistance.

Calibration Proceudre For Digital Multimeters (General) Page Navigation
  83    84    85    86    87  88  89    90    91    92    93  
img
TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
(All Manufacturers), Model M2625 - Continued
Table B-197. AC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V ac)
(V ac)
(V ac)
Min
Max
360 m @ 50 Hz
400 m
353.8  m
366.2  m
360 m @ 400 Hz
400 m
353.8  m
366.2  m
3.6 @ 50 Hz
4
3.566
3.634
3.6 @ 400 Hz
4
3.566
3.634
36 @ 50 Hz
40
35.66
36.34
36 @ 400 Hz
40
35.66
36.34
360 @ 50 Hz
400
356.6
363.4
360 @ 400 Hz
400
356.6
363.4
540 @ 50 Hz
600
529
551
540 @ 400 Hz
600
529
551
Table B-198. DC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A dc)
(A dc)
(A dc)
Min
Max
36  μ
40 μ
35.2  μ
36.8  μ
360  μ
400 μ
μ
μ
352
368
36  m
40 m
35.41 m
36.59 m
360  m
400 m
354.1  m
365.9  m
3.6
4
3.523
3.677
9
10
8.77
9.23
Table B-199. AC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(A ac)
(A ac)
Min
Max
36 μ @ 50 Hz
40 μ
34.8  μ
37.2  μ
36 μ @ 400 Hz
40 μ
34.8  μ
37.2  μ
360 μ @ 50 Hz
400 μ
μ
μ
348
372
360 μ @ 400 Hz
400 μ
μ
μ
348
372
36 m @ 50 Hz
40 m
35.23
36.77
36 m @ 400 Hz
40 m
35.23
36.77
360 m @ 50 Hz
400 m
352.3
367.7
360 m @ 400 Hz
400 m
352.3
367.7
3.6 @ 50 Hz
4
3.505
3.695
3.6 @ 400 Hz
4
3.505
3.695
9 @ 50 Hz
10
8.72
9.28
9 @ 400 Hz
10
8.72
9.28
B-78


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.