TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Wavetek, Model HD160 - Continued
Table B-192. DC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A dc)
(A dc)
(A dc)
Min
Max
9m
10 m
8.927 m
9.073 m
30 m
40 m
29.72 m
30.28 m
9
10
8.927
9.073
10
20
9.89
10.11
Table B-193. AC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(A ac)
(A ac)
Min
Max
9 m @ 50 Hz
10 m
8.837 m
9.163 m
9 m @ 400 Hz
10 m
8.837 m
9.163 m
9 m @ 1 kHz
10 m
8.837 m
9.163 m
30 m @ 50 Hz
40 m
29.42 m
30.58 m
30 m @ 400 Hz
40 m
29.42 m
30.58 m
30 m @ 1 kHz
40 m
29.42 m
30.58 m
9 @ 50 Hz
10
8.837
9.163
9 @ 400 Hz
10
8.837
9.163
9 @ 1 kHz
10
8.837
9.163
10 @ 50 Hz
20
9.79
10.21
10 @ 400 Hz
20
9.79
10.21
10 @ 1 kHz
20
9.79
10.21
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
1k
1k
-0.51% (994.9)
+0.51% (1.0051 k)
10 k
10 k
-0.51% (9.949 k)
+0.51% (10.051 k)
100 k
100 k
-0.51% (99.49 k)
+0.51% (100.51 k)
1M
1M
-0.51% (994.9 k)
+0.51% (1005.1 k)
10 M
10 M
-0.52% (9.948 M)
+0.51% (10.052 M)
19 M
40 M
-1.632% (18.69 M)
+1.632% (19.31 M)
B-76