TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Beckman Instruments, Model HD110B - Continued
Table B-180. AC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(A ac)
(A ac)
Min
Max
190 μ @ 45 Hz
200 μ
188.1 μ
191.9 μ
190 μ @ 1 kHz
200 μ
188.1 μ
191.9 μ
1.9 m @ 45 Hz
2m
1.881 m
1.919 m
1.9 m @ 1 kHz
2m
1.881 m
1.919 m
19 m @ 45 Hz
20 m
18.81 m
19.19 m
19 m @ 1 kHz
20 m
18.81 m
19.19 m
190 m @ 45 Hz
200 m
187.3 m
192.7 m
190 m @ 1 kHz
200 m
187.3 m
192.7 m
1.9 @ 45 Hz
2
1.873
1.927
1.9 @ 1 kHz
2
1.873
1.927
9 @ 45 Hz
10
8.82
9.18
9 @ 1 kHz
10
8.82
9.18
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
190
200
-0.7368% (188.6)
+0.7368% (191.4)
1.9 k
2k
-0.4211% (1.892 k)
+0.4211% (1.908 k)
19 k
20 k
-0.4211% (18.92 k)
+0.4211% (19.08 k)
190 k
200 k
-0.4211% (189.2 k)
+0.4211% (190.8 k)
1.9 M
2M
-0.7368% (1.886 M)
+0.7368% (1.914 M)
19 M
20 M
-1.4211% (18.73 M)
+1.4211% (19.27 M)
B-71