TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Beckman Instruments, Model HD110
Test instrument parameters
Performance specifications
DC VOLTAGE
Range: 0 to 1500 V dc
Accuracy: (0.25% rdg + 1 dgt)
AC VOLTAGE
Range: 0 to 1000 V ac
Accuracy: 45 Hz to 2 kHz, (0.75% rdg + 3 dgts);
2 kHz to 5 kHz, (1.5% rdg + 5 dgts);
5 kHz to 10 kHz, (2.5% rdg + 9 dgts)
DC CURRENT
Range: 0 to 10 A dc
Accuracy: 0 to 2 A dc rngs, (0.75% rdg + 1 dgt);
10 A dc rng, (1.5% rdg + 1 dgt)
Range: 0 to 10 A ac
AC CURRENT
Accuracy: 0 to 2 A ac rngs @ 45 Hz to 2 kHz, (1.5% rdg + 3 dgts);
10 A ac rng @ 45 to 400 Hz, (2% rdg + 3 dgts)
Range: 0 to 20 MΩ
Accuracy: 200 Ω to 2 MΩ rngs, (0.5% rdg + 1 dgt);
20 MΩ rng, (1.5% rdg + 1 dgt)
Table B-171. DC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V dc)
(V dc)
(V dc)
Min
Max
190 m
200 m
189.4 m
190.6 m
1.9
2
1.894
1.906
5
20
4.98
5.02
-5
20
-5.02
-4.98
10
20
9.96
10.04
-10
20
-10.04
-9.96
15
20
14.95
15.05
-15
20
-15.05
-14.95
19
20
18.94
19.06
-19
20
-19.06
-18.94
190
200
189.4
190.6
1000
1500
996
1004
B-66