TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Beckman Instruments, Model HD110B
Test instrument parameters
Performance specifications
DC VOLTAGE
Range: 0 to 1500 V dc
Accuracy: (0.1% rdg + 2 dgts)
AC VOLTAGE
Range: 0 to 1000 V ac
Accuracy: 200 mV ac to 200 V ac rngs @ 40 to 500 Hz, (0.8% rdg + 4 dgts);
200 mV ac to 20 V ac rngs @ 500 Hz to 5 kHz, (1.5% rdg + 5 dgts);
200 V ac rng @ 500 Hz to 2 kHz, (1.5% rdg + 5 dgts);
1000 V ac rng: @ 50 to 60 Hz, (0.8% rdg + 4 dgts);
@ 60 to 500 Hz, (1.5% rdg + 5 dgts)
DC CURRENT
Range: 0 to 10 A dc
Accuracy: 200 μA dc to 20 mA dc rngs, (0.5% rdg + 2 dgts);
200 mA dc and 2 A dc rngs, (1.0% rdg + 2 dgts);
10 A dc rng, (1.2% rdg + 2 dgts)
AC CURRENT
Range: 0 to 10 A ac, 45 Hz to 1 kHz
Accuracy: 200 μA ac to 20 mA ac rngs, (0.8% rdg + 4 dgts);
200 mA ac and 2 A ac rngs, (1.2% rdg + 4 dgts);
10 A ac rng, (1.5% rdg + 4 dgts)
Range: 0 to 20 MΩ
Accuracy: 200 Ω rng, (0.5% rdg + 4 dgts);
2 to 200 kΩ, (0.3% rdg + 2 dgts);
2 MΩ rng, (0.5% rdg + 4 dgts);
20 MΩ rng, (1.2% rdg + 4 dgts)
Table B-177. DC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V dc)
(V dc)
(V dc)
Min
Max
190 m
200 m
189.6 m
190.4 m
1.9
2
1.896
1.904
6
20
5.97
6.03
14
20
13.97
14.03
19
20
18.96
19.04
-19
20
-19.04
-18.96
190
200
189.6
190.4
1000
1500
997
1003
B-69