• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table B-164. Calibration Description.
Table B-169. Resistance

Calibration Proceudre For Digital Multimeters (General) Page Navigation
  69    70    71    72    73  74  75    76    77    78    79  
img
TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Beckman Instruments, Model HD100 - Continued
Table B-166. AC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V ac)
(V ac)
(V ac)
Min
Max
190 m @45 Hz
200 m
188.3  m
191.7  m
190 m @ 1 kHz
200 m
188.3  m
191.7  m
190 m @ 1.9 kHz
200 m
188.3  m
191.7  m
190 m @ 5 kHz
200 m
186.6  m
193.4  m
1.9 @45 Hz
2
1.883
1.917
1.9 @ 1 kHz
2
1.883
1.917
1.9 @ 1.9 kHz
2
1.883
1.917
1.9 @ 5 kHz
2
1.866
1.934
19 @45 Hz
20
18.83
19.17
19 @ 1 kHz
20
18.83
19.17
19 @ 1.9 kHz
20
18.83
19.17
19 @ 5 kHz
20
18.66
19.34
190 @45 Hz
200
188.3
191.7
190 @ 1 kHz
200
188.3
191.7
190 @ 1.9 kHz
200
188.3
191.7
190 @ 5 kHz
200
186.6
193.4
900 @45 Hz
1000
890
910
900 @ 1 kHz
1000
890
910
900 @ 1.9 kHz
1000
890
910
900 @ 5 kHz
1000
881
919
Table B-167. DC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A dc)
(A dc)
(A dc)
Min
Max
190  μ
200 μ
188.5  μ
191.5  μ
1.9 m
2m
1.885 m
1.915 m
19  m
20 m
18.85 m
19.15 m
190  m
200 m
188.5  m
191.5  m
1.9
2
1.885
1.915
Table B-168. AC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(V ac)
(A ac)
Min
Max
190 μ @ 45 Hz
200 μ
186.8  μ
193.2  μ
190 μ @ 2 kHz
200 μ
186.8  μ
193.2  μ
1.9 m @ 45 Hz
2m
1.868 m
1.932 m
1.9 m @ 2 kHz
2m
1.868 m
1.932 m
19 m @ 45 Hz
20 m
18.68 m
19.32 m
19 m @ 2 kHz
20 m
18.68 m
19.32 m
190 m @ 45 Hz
200 m
186.8  m
193.2  m
190 m @ 2 kHz
200 m
186.8  m
193.2  m
1.9 @ 45 Hz
2
1.868
1.932
1.9 @ 2 kHz
2
1.868
1.932
B-64


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.