TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Beckman Instruments, Model HD110B - Continued
Table B-178. AC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V ac)
(V ac)
(V ac)
Min
Max
190 m @ 40 Hz
200 m
188.1 m
191.9 m
190 m @ 500 Hz
200 m
188.1 m
191.9 m
190 m @ 550 Hz
200 m
186.6 m
193.4 m
190 m @ 5 kHz
200 m
186.6 m
193.4 m
1.9 @ 40 Hz
2
1.881
1.919
1.9 @ 500 Hz
2
1.881
1.919
1.9 @ 550 Hz
2
1.866
1.934
1.9 @ 5 kHz
2
1.866
1.934
19 @ 40 Hz
20
18.81
19.19
19 @ 500 Hz
20
18.81
19.19
19 @ 550 Hz
20
18.66
19.34
19 @ 5 kHz
20
18.66
19.34
190 @ 40 Hz
200
188.1
191.9
190 @ 500 Hz
200
188.1
191.9
190 @ 550 Hz
200
186.6
193.4
190 @ 2 kHz
200
186.6
193.4
900 @ 50 Hz
1000
889
911
900 @ 60 Hz
1000
889
911
900 @ 70 Hz
1000
881
919
900 @ 500 Hz
1000
881
919
Table B-179. DC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A dc)
(A dc)
(A dc)
Min
Max
190 μ
200 μ
188.8 μ
191.2 μ
1.9 m
2m
1.888 m
1.912 m
19 m
20 m
18.88 m
19.12 m
190 m
200 m
187.9 m
192.1 m
1.9
2
1.879
1.921
9
10
8.87
9.13
B-70