TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Beckman Instruments, Model HD110T - Continued
Table B-184. AC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V ac)
(V ac)
(V ac)
Min
Max
190 m @ 45 Hz
200 m
188.3 m
191.7 m
190 m @ 1.9 kHz
200 m
188.3 m
191.7 m
190 m @ 4.9 kHz
200 m
186.6 m
193.4 m
190 m @ 10 kHz
200 m
184.3 m
195.7 m
1.9 @ 45 Hz
2
1.883
1.917
1.9 @ 1.9 kHz
2
1.883
1.917
1.9 @ 4.9 kHz
2
1.866
1.934
1.9 @ 10 kHz
2
1.843
1.957
19 @ 45 Hz
20
18.83
19.17
19 @ 1.9 kHz
20
18.83
19.17
19 @ 4.9 kHz
20
18.66
19.34
19 @ 10 kHz
20
18.43
19.57
190 @ 45 Hz
200
188.3
191.7
190 @ 1.9 kHz
200
188.3
191.7
190 @ 4.9 kHz
200
186.6
193.4
190 @ 10 kHz
200
184.3
195.7
900 @ 45 Hz
1000
890
910
900 @ 1.9 kHz
1000
890
910
900 @ 4.9 kHz
1000
881
919
900 @ 10 kHz
1000
868
932
Table B-185. DC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A dc)
(A dc)
(A dc)
Min
Max
190 μ
200 μ
188.5 μ
191.5 μ
1.9 m
2m
1.885 m
1.915 m
19 m
20 m
18.85 m
19.15 m
190 m
200 m
188.5 m
191.5 m
1.9
2
1.885
1.915
9
10
8.85
9.15
B-73