TB 9-6625-1997-35
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
IF output
Frequency range: 4.3, 5.6, 5.625, 5.65, 10.0, and 115 MHz
11.5 MHz
Accuracy: 0.10%
2 Vrms to 1 Vrms
Level:
Accuracy: 2.5%
Distortion
<4%
Modulation deviation
Range:
10 to 40 kHz
Accuracy: 2 kHz on 10 kHz range
3 kHz on 20 kHz range
5 kHz on 40 kHz range
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer calibration
Standards Set AN/GSM-286, AN/GSM-287 and AN/GSM-705. Alternate items may be
used by the calibrating activity. The items selected must be verified to perform
satisfactorily prior to use and must bear evidence of current calibration. The equipment
must meet or exceed the minimum use specifications listed in table 2. The accuracies
listed in table 2 provide a four-to-one ratio between the standard and TI.
5. Accessories Required. The
accessories listed in
table 3 are
issued as
indicated in
paragraph 4 above and are
used in this
calibration procedure. When
necessary, these
items may be substituted by equivalent items unless specifically prohibited.
Table 2. Minimum Specifications of
Equipment Required
Minimum use
Manufacturer and model
Common name
specifications
(part number)
AUDIO ANALYZER
Range:
150 Hz to 1 kHz
Boonton, Model 1120-S/10
Accuracy: <1%
(MIS-35954/2)
Range:
950 Hz to 80 MHz
John Fluke, Model PM6681/656
Accuracy: 0.00094%
(PM6681/656)
MEASURING RECEIVER
Range:
8 to 45 kHz
Hewlett-Packard, Model 8902A
w/sensors,
Hewlett-Packard,
Accuracy: 1%
Model 11722A (11722A) and
Range:
7.5 mV to 1.25 V
11792A
(11792A),
and
(at 10 MHz)
microwave converter, model
Accuracy: 6.25%
11793A (11793A)
Range:
8 to 26 V dc
John Fluke, Model 8840A/AF-
05/09 (AN/GSM-64D)
Accuracy: 0.1%