TB 9-6625-1997-35
Test Instrument
Measuring Receiver
RF TUNING
BAND
RF OUTPUT
control for
SWITCH
switch setting
Min
Max
setting
(KUV)
(MHz)
(mV)
(mV)
B
30
62.5
46.9
78.1
125
93.75
156.25
250
187.5
312.5
500
375
625
A
20
62.5
46.9
78.1
125
93.75
156.25
250
187.5
312.5
500
375
625
C
45
62.5
46.9
78.1
125
93.75
156.25
250
187.5
312.5
500
375
625
D
65
62.5
46.9
78.1
125
93.75
156.25
250
187.5
312.5
500
375
625
(3) Repeat technique of (2) above for switch settings listed in table 6.
If
measuring receiver does not indicate within limits specified, perform b below.
b. Adjustments
(1) Set RF OUTPUT switch to 250 KUV and adjust LO-HI RF SET TO LINE
control for a 250 mV indication on measuring receiver.
NOTE
When TI protective cover is removed to make adjustments,
do not change setting of 115-230 LINE VOLTAGE
SELECTOR switch (located on bottom front center).
(2) Adjust R1 ((HI-RF METER CAL) located on lower right side) for a 0.55 or a
red-line indication on IF UV RF SET TO LINE meter (R).
a. Performance Check
TI IF output.
(2) Position controls as listed in (a) through (c) below:
CHANGE 1
9