TB 9-6625-1999-35
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the applicable sections before beginning
the calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. When indications specified in
paragraphs 8 through 15 are not within
tolerance,
perform the power supply check prior to making adjustments. After adjustments are made,
repeat paragraphs 8 through 15. Do not perform power supply check if all other
parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance checks where applicable.
a. Remove protective cover from TI only when necessary to make adjustments. Replace
cover after completing the adjustments.
c. Connect autotransformer to a 115 V ac
power source and adjust
controls for a
115 V ac output.
d. Set POWER switch to ON and allow at least 20 minutes for equipment to warm-up
and stabilize.
e. Set POWER switch to off (down) and wait 30 seconds.
f. If necessary, adjust mechanical zero adjust screws for a zero indication on R.M.S.
VOLTS and PERCENT MODULATION meters.
g. Set POWER switch to ON and allow 2 hours for equipment to warm-up and stabilize.