SECTION I. IDENTIFICATION AND DESCRIPTION
Equipment Setup

Calibration Procedure For Signal Generator, An/Urm-206 And Model 1608E-Y Page Navigation
  1  2  3    4    5    6    7  
img
. . . . . . . . .
TB 9-6625-2000-35
use and must bear evidence of current calibration. The equipment must meet or exceed the
minimum use specifications listed in table 2. The accuracies listed in table 2 provide a four-to-
one ratio between the standard and TI. Where the four-to-one ratio cannot be met, the actual
accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common usage
accessories, issued as indicated in paragraph 4 above and are not listed in the calibration
procedure.  The  following  peculiar  accessory  is  also  required  for  this  calibration:
SEMICONDUCTOR DEVICE (Coaxial Crystal Detector) Hewlett-Packard, Model 423AOPT03
(423AOPT03).
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
ATTENUATOR
Range: DC to 18 GHz
Weinschel, Model 9918-10dB (9918)
Accuracy:
0.5dB
AUTOTRANSFORMER
Range: 105 to 125 V ac
Ridge, Model 9020A (9020A)
1%
W10MT3AS3
Accuracy:
FREQUENCY COUNTER
Range: 6.5 to 11.5 GHz
Agilent, Model 5352B-001 (5352B-001)
Accuracy:
0.125%
Hewlett-Packard, Model 8902A
MEASURING RECEIVER
Frequency measurement:
w/sensors, Hewlett-Packard, Model
Range: 247.5 kHz to 80.8 MHz
11722A (11722A) and 11792A
Accuracy:
0.25%
(11792A), and microwave converter,
Power measurement:
model 11793A (11793A)
Range: +11 to -112 dBm
Accuracy:
0.25 dB;
-99 to +11 dBm
0.5 dB;
-112 to 108 dBm
MULTIMETER
Range: -1850 to +10 V dc
Fluke, Model 8840A/AF05
0.5%
(AN/GSM-64D)
Accuracy:
OSCILLOSCOPE
Range: 0 to 15 V pk 800 to 1200 Hz
(OS303/G)
3%
Accuracy:
SECTION III
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in this section are preparatory to the calibration process.
Personnel should become familiar with the entire bulletin before beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in the manufacturer's manual for this TI and
TM 11-6625-2948-14&P.
CHANGE 1
3