• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Equipment Setup
Display Accuracy

Calibration Procedure For Spectrum Analyzer, Model 3580A Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
. . . . .
TB 9-6625-2009-24
k. Repeat h and i above.
8. Frequency Calibration
a. Performance Check
(1) Connect equipment as shown in figure 1.
Figure 1. Frequency calibration - equipment setup.
(2) Position controls as listed in (a) through (h) below.
(a) AMPLITUDE MODE LOG 10 dB/DIV pushbutton pressed.
(b) FREQUENCY control to 00.0 Hz.
(c) RESOLUTION BANDWIDTH control to 30 Hz (or as required).
(d) START/CTR switch to CTR.
(e) FREQ SPAN/DIV control as required.
(f) SWEEP TIME/DIV control to .2 SEC.
(g) ZERO CAL control for maximum display at center line.
(h) INPUT SENSITIVITY switch to 0 (zero) dBv.
(3) Adjust calibrator output control to 0.3 V rms (0 (zero) dBm).
(4) Position FREQUENCY control as listed in table 3. Repeat (3) above and adjust
calibrator controls to peak signal pip on center of TI display.  Frequency counter will
indicate within limits specified.
6


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.